Bioimaging Systems formerly Atto BioscienceBD Bioscience formerly Atto Bioscience Site Map Contact Information BD Biosciences HomeSearch
BD™ CARV II takes you on a three-dimensional journey into a cell through time.
BD™ CARV II | Technical Specifications

Confocalscanner Nipkowspinning Disk (pinholes)
Disk scan rate 1000 scans per second
Pinholediameter 70 um
Spectraltransmission 360 nm– 700 nm
Z-resolution 0.5um(PSF); 100X PlanApo 1.4NA
Illuminationsource 120W Hg-Metal Halide (1200hr)
Internalexcitation changer Automated 8 position wheel (25 mm)
Internaldichroic changer Automated5 position wheel (25.7x36 mm)
Internalemission changer Automated 8 position wheel(25 mm)
Filtersets Provided DAPI,E-GFP, TEX RED; (Semrock Bright line series)
Operationmode Automatedconfocal, wide field, bright field
Observation Directconfocal binocular viewing or camera port
DetectorCompatibility CCDcamera: Sensicam EM, QE (Cooke) ; Coolsnap HQ, Cascade 512B (Photometrics);ORCA ER, AG (Hamamatsu) and more
MicroscopeCompatibility Mostinverted fluorescence microscopes with 100% camera port.
FRAP Aperture control - touch pad or RS232
Control Touch pador RS232
Softwaredrivers IP Lab(Scanalytics); Metamorph (UIC) RS232 command set also available
Size 11(w)x15.5 (L) x 6 (h) inches27.94(w)x 39.37 (L) x 15.24 cm
Weight 14.5 lbs/6.6Kg
Power 100-240VAC/12VDC