BD Pathway | Technical Specifications
Imaging Platform
- Proprietary, fully integrated optical design
- Motionless stage handles 96 or 384 well plates, and microscope slides
- X,Y resolution: 0.05 µm (mechanical)
- Z resolution: 0.1 µm (mechanical)
- 5 images per second speed
- Olympus 20X/0.75na UApo/340 0.17 mm working distance objective
- 4X, 10X, 40X, 60X objectives also available.
- Integrated temperature and CO2 control
Confocal Unit
- Disk sampling rate: 200 fps (actual image capture rate depends upon camera speed)
- Operation mode: Confocal, widefield, brightfield
- Pinhole configuration: Nipkow spinning disk
- Pinhole size: 70 µm
Laser-Free Illumination
- Excitation: 340 nm to near-IR
- Light source: Dual HBO 100W mercury arc lamp units (xenon lamps available)
- White-light LED for transmission microscopy (with optional liquid handling head)
Filter configuration (user specified)
- Excitation: 16 filter positions/5 dichroic mirror positions
- Emission: 8 filter positions/5 dichroic mirror positions
- Independent dichroic mirror selection
Detection/Observation
- Internal high resolution Hamamatsu ORCA ER CCD camera
- Binocular eyepiece for direct sample viewing (optional)
Liquid Handling
- Motorized head with single-position syringe (2 - 100 µl nominal capacity)
- Mixing capability (tituration)
Software
- Automated focus and image acquisition
- Automated region of interest (ROI) identification
- Interactive data and image integration
- Concurrent mulitple ratiometric dye kinetics (patented)
- Z sectioning and 3D rendering
- Novel heirarchical data classification algorithm
Computer (minimum specification)
- Pentium IV, 1.8 GHz running Windows 2000 including DVD burner
Physical Dimensions (inches/centimeters)
- Width: 43/109
- Height: 66/168
- Depth: 36/91 (including keyboard shelf)
|